GPT-201B is based on the differential pressure method and is professionally applicable to the determination of gas transmission rate as well as solubility coefficient, diffusion coefficient and permeability coefficient of plastic films, composite films, high barrier materials, sheeting, metal foils, rubber, tires and permeable membranes. Product Features
Gas transmission rate, solubility coefficient, diffusion coefficient, and permeability coefficient of the specimen could be obtained at one operation High precision and wide range of temperature and humidity control to meet different test conditions The instrument comes with three test modes: proportional mode, standard mode and time mode Test range could be extended based on user requirements to test materials with high permeability Test results could be easily obtained even at extreme conditions by data fitting function, which could work at any temperature The instrument could be used to test poisonous, inflammable, and explosive gases (customization required) Pneumatic sample clamp system assured the reliability of the test and makes the work high efficient. The instrument is controlled by the PLC control system and test process is automatic Reference film for fast calibration to ensure accurate and universal test data GMP three-level authority management Embedded microcomputer for test result printing. Professional software support real-time data display, data analysis and storage Standard USB port for convenient data transfer
Test Principle
The pre-conditioned specimen is mounted in the gas diffusion cell as to form a sealed barrier between two chambers. The lower-pressure chamber is firstly evacuated, followed by the evacuation of the entire cell. A flow of gas is thereafter introduced into the evacuated higher-pressure chamber and a constant pressure difference is generated between two chambers. The gas permeates through the specimen from the higher pressure side into the lower side. The gas permeability and other barrier properties of the specimen can be obtained by monitoring the pressure changes in the lower chamber.
Test Standards
This test instrument conforms to many national and international standards:
ISO 15105-1, ISO 2556, ASTM D1434, JIS K7126-1, GB/T 1038-2000, YBB 00082003